← All recalls
Medical Device

VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.

Recall date: 2026-10-28 · Source: FDA_DEVICE
Product
VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Units affected
4,369 units
Description
VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.
View official recall notice →