← All recalls
Medical Device

VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

Recall date: 2026-10-14 · Source: FDA_DEVICE
Brand / Firm
Product
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
Reason for recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Units affected
217
Description
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
View official recall notice →