Medical Device

SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model N

Recall date: 2026-09-30 · Source: FDA_DEVICE
Product
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782145 (3.0T), 5. 781270 (XR and 3.0T), 6. 782113 (XR and 3.0T), 7. 782129 (XR and 3.0T), 8. 782118 (Ingenia Elition X), 9. 782144 (Ingenia Elition X);
Reason for recall
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Units affected
965
Description
SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, SmartPath to Ingenia Elition X; Model Numbers (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782145 (3.0T), 5. 781270 (XR and 3.0T), 6. 782113 (XR and 3.0T), 7. 782129 (XR and 3.0T), 8. 782118 (Ingenia Elition X), 9. 782144 (Ingenia Elition X);