← All recalls
Medical Device

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117

Recall date: 2026-10-21 · Source: FDA_DEVICE
Brand / Firm
Product
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117
Reason for recall
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Units affected
106758 units
Description
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117
View official recall notice →