Medical DeviceView official recall notice →
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been d
Brand / Firm
Product
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
Reason for recall
Devices which did not undergo thermoforming could deform and lose performance.
Units affected
1490 units (OUS only)