Medical DeviceView official recall notice →
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro So
Brand / Firm
Product
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Reason for recall
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Units affected
167 units
Description
Philips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key