Medical Device

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Cove

Recall date: 2026-09-02 · Source: FDA_DEVICE
Brand / Firm
Product
ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000
Reason for recall
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Units affected
6 units
Description
ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000