Optima XR240amx, Mobile Digital Radiographic System
Brand / Firm
Product
Optima XR240amx, Mobile Digital Radiographic System
Reason for recall
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Units affected
24 units
Description
Optima XR240amx, Mobile Digital Radiographic System