Medical Device

Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer S

Recall date: 2026-05-27 · Source: FDA_DEVICE
Brand / Firm
Product
Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960
Reason for recall
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
Units affected
20,506
Description
Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960