Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova
Brand / Firm
Product
Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Intera 1.5T Omni/Stellar (781104), 7. Intera Achieva 1.5T Pulsar (781171);
Reason for recall
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Units affected
82