← All recalls
Medical Device

MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND103

Recall date: 2026-06-24 · Source: FDA_DEVICE
Product
MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Units affected
8327756 units
Description
MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
View official recall notice →