← All recalls
Medical Device

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU

Recall date: 2026-06-24 · Source: FDA_DEVICE
Product
MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Units affected
1248 units
Description
MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.
View official recall notice →