← All recalls
Medical Device

MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C

Recall date: 2026-06-24 · Source: FDA_DEVICE
Product
MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C
Reason for recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Units affected
8260 units
Description
MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C
View official recall notice →