Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breat
Brand / Firm
Product
Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit
Reason for recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Units affected
20 units
Description
Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.