Medical DeviceView official recall notice →
Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MG
Brand / Firm
Product
Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP TRAY, Model Number:DYNJRA1979; 5) CPNB SETUP TRAY, Model Number:DYNJRA1979A; 6) CPNB SETUP TRAY, Model Number:DYNJRA1979B; 7) VARITHENA TRAY 2, Model Number:DYNJRA2608; 8) CPNB SETUP TRAY, Model Number:SAMPA0108
Reason for recall
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Units affected
1774 units
Description
Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number:DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number:DYNJ56436B; 3) SCC RF PAIN MGMT, Model Number:DYNJ56436C; 4) CPNB SETUP TRAY, Model Number:DYNJRA1979; 5) CPNB SETUP TRAY, Model Number:DYNJRA1979A; 6) CPNB SETUP TRAY, Model Number:DYNJRA1979B; 7) VARITHENA TRAY 2, Model Number:DYNJRA2608; 8) CPNB SETUP TRAY, Model Number:SAMPA0108