Product
Medline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, Model Number: DYKMBNDL199A
Reason for recall
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Units affected
104 kits
Description
Medline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2) KIT ANESTHESIA L&D A- LINE SET, Model Number: DYKMBNDL199A
View official recall notice →