LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors

Recall date: 2026-04-29 · Source: FDA_DEVICE
Product
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)
Reason for recall
Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Units affected
136
Description
LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)