Medical DeviceView official recall notice →
LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.
Brand / Firm
Product
LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.
Reason for recall
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
Units affected
321 units
Description
LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.