Medical Device

Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number:

Recall date: 2026-10-07 · Source: FDA_DEVICE
Product
Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual
Reason for recall
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Units affected
61,858
Description
Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual
View official recall notice →