GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

Recall date: 2026-08-12 · Source: FDA_DEVICE
Product
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Reason for recall
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Units affected
163 units
Description
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
View official recall notice →