Medical DeviceView official recall notice →
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320
Brand / Firm
Product
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320
Reason for recall
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Units affected
632
Description
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320