Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167

Recall date: 2026-03-04 · Source: FDA_DEVICE
Product
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Reason for recall
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Units affected
150
Description
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167