Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Recall date: 2026-02-11 · Source: FDA_DEVICE
Product
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
Reason for recall
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Units affected
3050 units
Description
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;