Medical DeviceView official recall notice →
B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR com
Brand / Firm
Product
B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor
Reason for recall
Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
Units affected
5,524
Description
B¿R¿A¿H¿M¿S KRYPTOR compact REACT, REF: 89986, used with B¿R¿A¿H¿M¿S PlGF Plus KRYPTOR, REF: 859075N, and run on the B¿R¿A¿H¿M¿S KRYPTOR compact PLUS analyzers. in vitro diagnostic test for Placental Growth Factor