← All recalls
Medical Device

ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.

Recall date: 2026-04-15 · Source: FDA_DEVICE
Product
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
Reason for recall
During patient examinations, the system may sporadically display a major increase of the x-ray dose applied to the patient. Despite the reported dose showing significantly higher, the actual applied x-ray dose levels are in accordance with the system settings and imaging conditions, as well as the applicable regulatory requirements.
Units affected
U.S. 758, OUS 2077
Description
ARTIS pheno and ARTIS Icono biplane, floor and ceiling. Model Numbers 10849000, 11327600, 11328100, 11327700.
View official recall notice →